Expression of the uncertainty of measurement in the calibration of thermometers Part II: Thermocouples
dc.contributor.author | Lira, I | |
dc.contributor.author | Santos, PR | |
dc.date.accessioned | 2025-01-21T01:32:16Z | |
dc.date.available | 2025-01-21T01:32:16Z | |
dc.date.issued | 1999 | |
dc.description.abstract | A model is proposed to evaluate the uncertainty of measurement in the calibration of thermocouples. The model applies to thermocouple types standardized in accordance with electromotive force-temperature reference functions. The difference between the calibration uncertainty and the user's uncertainty is established. Equations are illustrated through an example involving a type S thermocouple calibrated at four fixed points of the International Temperature Scale of 1990 (ITS-90). The example demonstrates the strong dependence of the calibration uncertainty on the selected form of the deviation function. | |
dc.fuente.origen | WOS | |
dc.identifier.issn | 0026-1394 | |
dc.identifier.uri | https://repositorio.uc.cl/handle/11534/97235 | |
dc.identifier.wosid | WOS:000085325700003 | |
dc.issue.numero | 5 | |
dc.language.iso | en | |
dc.pagina.final | 419 | |
dc.pagina.inicio | 415 | |
dc.revista | Metrologia | |
dc.rights | acceso restringido | |
dc.subject.ods | 13 Climate Action | |
dc.subject.odspa | 13 Acción por el clima | |
dc.title | Expression of the uncertainty of measurement in the calibration of thermometers Part II: Thermocouples | |
dc.type | artículo | |
dc.volumen | 36 | |
sipa.index | WOS | |
sipa.trazabilidad | WOS;2025-01-12 |